| Test & Calibration | Air Flow & Environmental | Chart Recorders | |||
|
Highly Precise (0,01%) Pressure Transmitters Series 33 X / 35 X
Digital
Output of Transmitter This high precision of 0,01 %FS is available as an option (the standard Series 33 X has an accuracy of 0,05 %FS). These Series are based on the stable, floating piezoresisitive transducer and the newly developed XEMICS micro-processor with integrated 16 bit A/D converter. Temperature dependencies and non-linearities of the sensor are mathematically compensated. With the READ30 software and the KELLER cable K-107, the calculated pressure can be displayed on a Palmtop, Laptop or PC. The READ 30 software also allows the recording of pressure signals and the graphic display on the PC. Up to 128 transmitters can be hooked together to a Bus-system. Transmitter with Analog Output Integrated in the XEMICS processor is a D/A converter of 16 bit for analog signal outputs of 4 20 mA or 0 10 V. The output rate is 400 Hz. The accuracy is diminished by this converting process by 0,05 %FS. The digital output is available on all transmitters with analog output. Programming With the KELLER software READ 30 and PROG30, a RS485 converter (i.e. K102 or K107 from KELLER) and a PC, the pressure can be displayed, the units changed, a new gain or zero set. The analog output can be set to any range within the compensated range. |
|
![]() |
![]() |
|
Series
33 X
Male Pressure Port G1/4" |
Series
35 X
Pressure Port G1/2'', Flush Diaphragm |
| Pressure Ranges | 0,8...1000 bar (abs./rel.) |
| Output Signal |
RS485
(digital) 4...20 mA (2-wire) 0...10 V (3-wire) |
|
Accuracy,
Error Band (10 40 °C) |
typ. 0,05 %FS
(digital) typ. 0,15 %FS (2-wire) typ. 0,1 %FS (3-wire) |
|
Optional:
Precision * |
typ. 0,01 %FS (digital) |
| Long Term Stability typ. |
Gauges:
1 mbar or 0,05 %FS Absolute: 0,5 mbar or 0,025 %FS (10 40 °C) |
| Resolution | 0,002 %FS |
|
*
|
Only
for Series 33 X and for ranges higher than 10 bar. Accuracy and Precision Accuracy is an absolute term, Precision a relative term. Dead weight testers are primary standards for pressure, where the pressure is defined by the primary values of mass, length and time. Highest class primary standards in national laboratories indicate the uncertainty of their pressure references with 70 to 90 ppM or close to 0,01%. Commercial dead weight testers as used in our facilities to calibrate the transmitters indicate an uncertainty or accuracy of 0,025 %. Below these levels, KELLER use the expression Precision as the ability of a pressure transmitter to be at each pressure point within 0.01 %FS relative to these commercial standards. The transmitters full-scale output can be set up to match any standard of your choice by correcting the gain with the PROG30 software. |
| Test & Calibration | Air Flow & Environmental | Chart Recorders | |||
![]() |
|
Or call us now on |
|
UK / Europe Office Tel +44 (0)8700 43 40 40 Fax +44 (0)8700 43 40 45 info@omniinstruments.co.uk East Kingsway Business Centre
Mid
Craigie Trading Estate,
Mid Craigie Rd,
Dundee,
DD4 7RH, Scotland, UK |
USA / Canada Office
Tel +1-866-849-3441
Fax +1 866 628 8055
info@omniinstruments.net
www.omniinstruments.net Tel +61 (0)282 442 363 Fax +61 (0)294 751 278 info@omniinstruments.com.au www.omniinstruments.com.au |
|||||
|
|
|